Single Analyte Ozone Monitor

Posted by Nathan on April 15, 2010 under Commercial Ozone Usage, Ozone Products, Wastewater Treatment | Be the First to Comment

We now offer an alternative for calibrating larger dissolved monitors or to test dissolved ozone levels in a lab setting.  The SAM-2 and the SAM-3 monitors are inexpensive dissolved ozone monitors that are both compact and reliable.

The SAM-2 has a range of 0.15-0.75 PPM, making it perfect for lab settings. No activator solution is required, which helps to cut costs. This monitor reads the ampoule that needs to be tested, and then will read a blank reference ampoule that is provided. If your ozone levels exceed its range, that is okay. Just dilute a fresh sample and retry the test! When you multiply the reading by the dilution factor you will have an accurate result. It is that simple.


SAM-3 Single Analyte Ozone MeterThe SAM-3 has a range of 0.20-3.00 PPM which makes it more appropriate for calibrating dissolved ozone monitors that are constantly in use (for example, an AS-500 on an ozone injection system). Using this analyte meter to calibrate your monitor monthly will help ensure accurate readings. This unit is also very simple to use – just zero it using the provided blank ampoule, and it is ready to test. This only needs to be done once (not between every test), which will speed up the testing process. After the unit is zeroed, you can test any number of ampoules.

You can get more information on these monitors by visiting the SAM-2 or SAM-3 on our website.