High Concentration UV Ozone Analyzer
High Range Ozone Analyzer
The UV-106H is designed to measure high concentrations of ozone gas. Ozone concentrations up up to 20% by weight can be measured with the UV-106H.
Versatile, easy to use, UV-106H is a great tool to assist in determine the actual ozone production of your ozone generator. Use in-line or for periodic spot checking.
Features
- 0.01 wt% Accuracy
- Reads High Range Ozone Levels
- Small, Compact Size
- Data Logging Capabilities
| Product ID: | UV-106H Rental | |
| In-stock: | yes |
Rental Rates:
| 7-day: | $495.00 | |
| 30-day: | $995.00 |
Specifications
Measurement Principle: | UV Absorbance |
Linear Dynamic Range: | 0-20 wt% |
Resolution: | 0.01 wt% |
Precision (1o for 10-s average; aka rms noise): | 0.01 wt% or 2 % of reading |
Accuracy: | 0.01 wt% or 2 % of reading |
Calibration: | Iodometric Titration |
Limit of Detection (2o): | 0.02 wt% |
Flow Rate (nominal): | 1 L/min |
Flow Rate Requirement: | >0.6 L/min |
Baseline Drift: | <0.02%/day, <0.1%/year |
Sensitivity Drift: | <1%/day, <3%/year |
Measurement Time and Frequency: | 10 s, 0.1 Hz |
Response Time, 100% of Step Change: | 20 s 2 points |
Averaging Times: | 10 s, 1 min, 5 min, 1 hr |
Ozone Units: | mol %, wt % |
Pressure Units: | mbar, torr |
Temperature Units: | °C, K |
T and P Corrected: | Yes |
Operating Temperature Range: | 0 to 50 °C |
Operating Altitude: | ~0-5 km |
Data Logger Capacity: | 14,336 lines |
Serial Output, RS232: | Yes |
USB Output: | Yes |
Data Transfer Baud Rates: | 2400, 4800, 19200 |
Analog Voltage Output: | 0-2.5 V |
Analog Current Output: | 4-20 mA |
Output Ranges: | User Defined Scaling Factor in Menu |
Dimensions: | 4 x 7.5 x 8.5-in (HxWxD) 10 x 19 x 22-cm |
Weight: | 4-lb (2.0 kg) |