S-500L Rental Support
|
|
Support Related Frequently Asked Questions
Q.How does “Control Hi/Low” differ from “Alarm Hi/Lo”? A.Control Hi/Low permits an internal circuit to close based on a high or low ozone level present. Alarm Hi/Low is for triggering the audible alarm on the unit as well as external alarms (Pins 4 & 5). Sometimes, people will want an audible alarm with an external signal sent out before any action is taken with the control set point (Pin 3). Q.Does the S-500 show ozone levels on a computer screen in real time? A.Yes. Ozone levels will plot to the computer in real time. This data is also available for viewing with included software and an CSV file for download. Q.Can the S-500 be connected to a USB port? A.The S-500 comes with an RS-232 cable to connect the monitor to a computer. A male RS-232 to USB adapter is required. This item can be purchased from Ozone Solutions for $19.95. The part# is RS-232-USB. Q.How well do the ozone sensors work with caprolactam vapors? A.The sensors are unlikely to read ozone in the presence of caprolactam vapor. This is because there may be cross interference with the caprolactam, and the presence of caprolactam is likely to drive the reading down faster than ozone will drive the reading upwards. Therefore it is necessary to position the sensor so that ozone is read but caprolactam is not present.
|