Inexpensive Dissolved Ozone Monitor
Inline Dissolved O3 sensor
The DOM-1 is an inexpensive inline dissolved ozone sensor. 0 - 2.00 PPM range with 0.01 PPM resolution. Low Cost, Rugged Dissolved Ozone Monitor.
Features
- Can handle high pressure water streams
- Specific to Ozone
- Low cost and easily replaceable parts
- Set-point relay for alarm or system control
- 4-20 mA output to PLC or other remote readouts and controllers
- Accurate ozone levels with even discolored or dirty water
Related Files
|
|
DOM-1 System Placement |
| 44.5K | |
|
|
DOM-1 Drain Placement |
| 59.7K |
| Product ID: | DOM-1 | |
| Lead Time: | 3 weeks | |
| Price: | $2,150.00 |
Calibration Kit
Dissolved Ozone Monitor must be calibrated on site and in your water. If you do not currently have an ozone test kit we recommend purchasing this colorimetric test kit:
| Part Number | Price | Buy Online | |
|---|---|---|---|
| K-7402 | $74.00 |
Specifications
Accuracy: | Better than 20%; 10% typical |
Readout: | 0.03 - 2.00 ppm in 0.01 ppm increments |
Outputs: | NO/NC contacts & 4-20mA |
Relay Ratings: | 5 Amps at 250V AC |
Lowest reading: | 0.03 ppm |
Protection Specifications: | IP 52. Enclosure is polycarbonate. Front cover is clear polycarbonate |
Warm-Up: | 1 hour |
Response Time: | 20-30 seconds |
Supply Water Pressure: | 15 - 90 PSI |
Water Temperature Range: | 68 to 86 F (20-30 deg C)--unless re-calibrated for a specific temperature |
Power Source: | 12 VDC 800 mA [AC Adapter included] |
Requirements: | |
Water feed at > 14 psi (1 bar) | |
0.15 liters/minute (3 gallons/hour) flow to drain or reclaim | |
Calibration: | Sensor must be calibrated monthly to a coloremetric standard on a weekly basis for best accuracy. |
Note: | DOM-1 can measure ozone accurately up to 150 PPM when used with a specially calibrated UV-100 Ozone Analyzer! - Call for details |
Warranty: | 1 year |