Model 400E Photometric O3 Analyzer

ambient ozone analyzer

High accuracy photometric ozone analyzer. This unit is ideal for laboratory or research applications where accuracy is a necessity.

Features

  • 0-100 ppb to 0-10 ppm ranges, user-selectable
  • Dual ranges and auto-ranging
  • Microprocessor-based with simple, multitasking, menu-driven software
  • Test and diagnostic functions with continuous self-check and alarms
  • Temperature and pressure compensation, auto-zero adjustment
  • Internal data logger with recording intervals from 1 min to 365 day multiple averages for up to 1 mi
  • Four analog outputs, two serial ports and APIcom Windows-based operating software

Related Files

400E Info Sheet 400E Info Sheet
172.8K
Product ID: API-400E
Lead Time:4 weeks

Specifications

Reporting Units:ppm, ppb, mg/m3, micro g/m3 / ug/m3
Zero noise:< 0.3 ppb (RMS)
Span noise:< 0.5% of reading (RMS) (above 100 ppb)
Lower detectable limit (LDL):< 0.6 ppb (RMS) (2x noise as per US EPA)
Zero drift:< 1 ppb/24 hours, < 1 ppb/7 days
Span drift:< 1% reading/7 days
Lag time:< 10 seconds
Rise and fall time:< 20 seconds to 95%
Linearity:1% full scale
Precision:0.5% of reading above 50 ppb
Pneumatics:
Internal diaphragm-type pump
800 cm3/min +/- 10%
47 mm diameter particulate filter, front-panel accessible
Operating Range:5-40 ºC (with EPA equivalency)
Dimensions (HxWxD):7" (178 mm) x 17" (432 mm) x 23.5" (597 mm)
Weight:28 lbs. (13 kg), 31 lbs. (14 kg) with IZS installed
Power:100-130 or 220-240 V, both at 50/60Hz, 250 W
Approvals:US EPA designation EQOA-0992-087 CE Mark Compliance
Outputs:Analog 0.1V/1V/5V/10V full scale, user selectable, -5 to +10% offset 8 digital status outputs, 6 inputs
Digital: COM1: RS-232 Serial Port (DB-9M connector) (incl. cable) COM2: RS-232 or RS-485 (DB-9F connector)